Recommended Equipment


The LEADseeker™ Multimodality Imaging System CCD imager can measure SPA Imaging assays, Cy Dye fluors, HitHunter™ chemiluminescent and fluorescent end point Enzyme Fragment Complementation Assays as well as a wide range of fluorescent and luminescent tags in 96, 384 and 1536 well microplates. Appropriate filter sets are available for all these end points.

SPA Scintillation Beads can be counted on a range of scintillation counters. For counters fitted with spectrum analysis packages, the suitable window opening can be determined and the instrument set accordingly.

For further information regarding programming and calibration of counters, please contact your instrument manufacturer.


Counter type
Isotope and SPA bead type
3H
125I
33P
35S
14C
PVT
YSi
PVT
YSi
PVT
YSi
PVT
YSi
PVT
YSi
Packard TopCount
1.5-35
0-50
1.5-100
0-100
2.9-256
2.9-256
2.9-256
2.9-256
2.9-256
2.9-256
Wallac MicroBeta
5-360
5-500
5-530
5-650
5-650
5-750
5-600
5-700
5-650
5-750
Hidex Chameleon*
20-120
20-120
20-900
20-900
20-600
20-600
N.D
N.D
N.D
N.D
Beckman LS7800
H# is set at 0 and the window is wide open from 10 to 999
Packard Tricarb 460
Window open from 0 to 999
LKB 1209/1215 RackBeta
Window open from 5 to 999

Instruments settings for SPA Scintillation beads - MicroBetaTM
PVT
YSi
isotope
window setting
efficiency (%)
window setting
efficiency (%)
3H
5 - 320
55
5 - 500
80
33P
5 - 600
60
5 - 750
37
35S
5 - 600
57
5 - 700
52
125I
5 - 530
49
5 - 650
87

Instruments settings for SPA Scintillation beads - TopCountTM
PVT
YSi
isotope
window setting
efficiency (%)
window setting
efficiency (%)
3H
1.5 - 35
34
0 - 50
69
33P
1.6 - 150
47
2.9 - 256
30
35S
1.5 - 150
40
2.9 - 256
43
125I
1.5 - 100
33
0 - 100
75

Instrumentation Settings
TopCount
Version 4 software and above
scintillator – liquid/plastic
energy range – low
mode – normal
scintillator - glass energy range - low mode - high sensitivity
TopCount
Version 3 software and below
scintillator – liquid
energy range – low
mode – high
scintillator - solid energy range - low mode - normal
Other instruments e.g. RackBeta or Tri-Carb Liquid Scintillation Analyzer
Applicable software
wide open
wide open




New to Scintillation Proximity Assay (SPA)? Get started with the basics

Use theSPA selection guide to see the entire product range

Understand the safety issues involved in using this technology

Visit the literature section and browse from over 800 references

Consult the technical support and customer service section for more support